Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI method.
- Source :
-
Journal of Cardiovascular Magnetic Resonance (BioMed Central) . 2013, Vol. 15 Issue 1, p63-70. 8p. 1 Color Photograph, 7 Graphs. - Publication Year :
- 2013
-
Abstract
- Background: Myocardial T1-mapping methods such as MOLLI use SSFP readout and are prone to frequency-dependent error in T1-measurement. A significant error in T1 may result at relatively small off-resonance frequencies that are well within the region without banding artifacts. Methods: The sensitivity of T1-estimates based on the SSFP based MOLLI sequence to errors in center frequency are calculated by means of a Bloch simulation and validated by phantom measurements. Typical off-resonance errors following local cardiac shimming are determined by field mapping at both 1.5 and 3.0T. In vivo examples demonstrate the artifactual appearance of T1-maps in the presence of off-resonance variation. Results: Off-resonance varied 61.8 ± 15.5 Hz (mean ± SD, n = 18) across the heart at 1.5T and 125.0 ± 40.6 Hz (mean ± SD, n = 18) at 3.0T. For T1 = 1000 ms, the variation in T1 due to off-resonance variation was approximately 20 ms at 62 Hz, and > 50 ms at 125 Hz. Conclusions: Regional variations due to the inability to completely shim the B0-field variation around the heart appear as regional variation in T1, which is artifactual. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 1532429X
- Volume :
- 15
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- Journal of Cardiovascular Magnetic Resonance (BioMed Central)
- Publication Type :
- Academic Journal
- Accession number :
- 89719142
- Full Text :
- https://doi.org/10.1186/1532-429X-15-63