Abstract: This paper deals with the direct measurement of total (fixed plus removable) surface activity in the presence of interfering radiation fields. Two methods based on Monte Carlo simulations are used: one for a Geiger–Muller (GM) ionisation probe and the other for sodium iodide (NaI) detector with lead collimators; equations for the most general case and the geometry models for Monte Carlo simulation of both (GM and NaI) detectors are employed. Finally, an example of application is discussed. [Copyright &y& Elsevier]