Back to Search Start Over

Development of CdTe pixel detector for synchrotron radiation experiments.

Authors :
Hirono, T.
Toyokawa, H.
Furukawa, Y.
Kawase, M.
Ohata, T.
Wu, S.
Ikeda, H.
Sato, G.
Takahashi, T.
Watanabe, S.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Dec2013, Vol. 731, p64-67. 4p.
Publication Year :
2013

Abstract

Abstract: This study describes the development of a CdTe pixel detector for high-energy X-ray diffraction experiments at synchrotron radiation facilities. We have developed three prototypes. The third prototype detector (SP8-02B) was designed with a pixel size of and a matrix of 20×50 pixels. The sensor was bonded to an application-specific integrated circuit with a preamplifier, shaper, window-type discriminator, and a 20-bit counter for each pixel. The SP8-02B detector improved the uniformity between pixels in one chip in comparison with the previous prototypes. The long-term stability was measured at various temperatures. A moderate cooling achieved good stability that was not affected by the polarization effect. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
01689002
Volume :
731
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
91970090
Full Text :
https://doi.org/10.1016/j.nima.2013.06.049