Development of CdTe pixel detector for synchrotron radiation experiments.
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Dec2013, Vol. 731, p64-67. 4p. - Publication Year :
- 2013
-
Abstract
- Abstract: This study describes the development of a CdTe pixel detector for high-energy X-ray diffraction experiments at synchrotron radiation facilities. We have developed three prototypes. The third prototype detector (SP8-02B) was designed with a pixel size of and a matrix of 20×50 pixels. The sensor was bonded to an application-specific integrated circuit with a preamplifier, shaper, window-type discriminator, and a 20-bit counter for each pixel. The SP8-02B detector improved the uniformity between pixels in one chip in comparison with the previous prototypes. The long-term stability was measured at various temperatures. A moderate cooling achieved good stability that was not affected by the polarization effect. [Copyright &y& Elsevier]