32P Brachytherapy Conformal Source Model RIC-100 for High-Dose-Rate Treatment of Superficial Disease: Monte Carlo Calculations, Diode Measurements, and Clinical Implementation.

Authors :
Cohen, Gil'ad N.
Munro, John J.
Kirov, Assen
Losasso, Thomas
Yamada, Yoshiya
Williamson, Matthew
Dauer, Lawrence T.
Zaider, Marco
Source :
International Journal of Radiation Oncology, Biology, Physics. Mar2014, Vol. 88 Issue 3, p746-752. 7p.
Publication Year :
2014

Abstract

Purpose: A novel 32P brachytherapy source has been in use at our institution intraoperatively for temporary radiation therapy of the spinal dura and other localized tumors. We describe the dosimetry and clinical implementation of the source. Methods and Materials: Dosimetric evaluation for the source was done with a complete set of MCNP5 Monte Carlo calculations preceding clinical implementation. In addition, the depth dose curve and dose rate were measured by use of an electron field diode to verify the Monte Carlo calculations. Calibration procedures using the diode in a custom-designed phantom to provide an absolute dose calibration and to check dose uniformity across the source area for each source before treatment were established. Results: Good agreement was established between the Monte Carlo calculations and diode measurements. Quality assurance measurements results are provided for about 100 sources used to date. Clinical source calibrations were usually within 10% of manufacturer specifications. Procedures for safe handling of the source are described. Discussion: Clinical considerations for using the source are discussed. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
03603016
Volume :
88
Issue :
3
Database :
Academic Search Index
Journal :
International Journal of Radiation Oncology, Biology, Physics
Publication Type :
Academic Journal
Accession number :
94404542
Full Text :
https://doi.org/10.1016/j.ijrobp.2013.11.006