Effect of Scan Quality on Diagnostic Accuracy of Spectral-Domain Optical Coherence Tomography in Glaucoma.

Authors :
RAO, HARSHA L.
ADDEPALLI, UDAY K.
YADAV, RAVI K.
SENTHIL, SIRISHA
CHOUDHARI, NIKHIL S.
GARUDADRI, CHANDRA S.
Source :
American Journal of Ophthalmology. Mar2014, Vol. 157 Issue 3, p719-726. 10p.
Publication Year :
2014

Abstract

• PURPOSE: To evaluate the effect of scan quality on the diagnostic accuracies of optic nerve head (ONH), retinal nerve fiber layer (RNFL), and ganglion cell complex (GCC) parameters of spectral-domain optical coherence tomography (SD OCT) in glaucoma. • DESIGN: Cross-sectional study. • METHODS: Two hundred fifty-two eyes of 183 control subjects (mean deviation [MD]:L1.84 dB) and 207 eyes of 159 glaucoma patients (MD: L7.31 dB) underwent ONH, RNFL, and GCC scanning with SD OCT. Scan quality of SD OCT images was based on signal strength index (SSI) values. Influence of SSI on diagnostic accuracy of SD OCT was evaluated by receiver operating characteristic (ROC) regression. • RESULTS: Diagnostic accuracies of all SD OCT parameters were better when the SSI values were higher. This effect was statistically significant (P < .05) for ONH and RNFL but not for GCC parameters. In mild glaucoma (MD of L5 dB), area under ROC curve (AUC) for rim area, average RNFL thickness, and average GCC thickness parameters improved from 0.651, 0.678, and 0.726, respectively, at an SSI value of 30 to 0.873, 0.962, and 0.886, respectively, at an SSI of 70. AUCs of the same parameters in advanced glaucoma (MD of L15 dB) improved from 0.747, 0.890, and 0.873, respectively, at an SSI value of 30 to 0.922, 0.994, and 0.959, respectively, at an SSI of 70. • CONCLUSION: Diagnostic accuracies of SD OCT parameters in glaucoma were significantly influenced by the scan quality even when the SSI values were within the manufacturer-recommended limits. These results should be considered while interpreting the SD OCT scans for glaucoma. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00029394
Volume :
157
Issue :
3
Database :
Academic Search Index
Journal :
American Journal of Ophthalmology
Publication Type :
Academic Journal
Accession number :
94512628
Full Text :