Back to Search Start Over

Reconstruction Method for Nonconcyclic Dual-Source Circular Cone-Beam CT with a Large Field of View.

Authors :
Ming Chen
Gang Li
Weiwei Qi
Jing Zou
Yong-guo Zheng
Source :
Mathematical Problems in Engineering. 2013, p1-7. 7p.
Publication Year :
2013

Abstract

In industrial computed tomography (CT), it is often required to inspect large objects whose size is beyond a reconstructed field of view (FOV). Somemultiscanmodes have been developed to acquire the complete CT projection data for a larger object using small panel detectors. In this paper, we give a non-concyclic dual-source circular cone-beam scanning geometry based on the idea of multiscan modes and propose a backprojection-filtration-based (BPF) reconstruction algorithm without data rebinning. Since the FOV calculated according to this nonconcyclic dual-source circular CT scanning geometry is larger than cardiac dual-source CT scanning geometry, our method can reconstruct larger horizontal slices (i.e., the slices perpendicular to rotation axis) than cardiac dual-source CT. The quality of CT images is expected to be superior to those obtained using larger panel detectors. The simulation results have indicated that CT images obtained by the proposed method are satisfying. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
1024123X
Database :
Academic Search Index
Journal :
Mathematical Problems in Engineering
Publication Type :
Academic Journal
Accession number :
94813373
Full Text :
https://doi.org/10.1155/2013/526247