Assessment of the dosimetric accuracies of CATPhan 504 and CIRS 062 using kV-CBCT for performing direct calculations.

Authors :
Annkah, James Kwame
Rosenberg, Ivan
Hindocha, Naina
Moinuddin, Syed Ali
Ricketts, Kate
Adeyemi, Abiodun
Royle, Gary
Source :
Journal of Medical Physics. Jul-Sep2014, Vol. 39 Issue 3, p133-141. 9p.
Publication Year :
2014

Abstract

The dosimetric accuracies of CATPhan 504 and CIRS 062 have been evaluated using the kV-CBCT of Varian TrueBeam linac and Eclipse TPS. The assessment was done using the kV-CBCT as a standalone tool for dosimetric calculations towards Adaptive replanning. Dosimetric calculations were made without altering the HU-ED curves of the planning computed tomography (CT) scanner that is used by the Eclipse TPS. All computations were done using the images and dataset from kV-CBCT while maintaining the HU-ED calibration curve of the planning CT (pCT), assuming pCT was used for the initial treatment plan. Results showed that the CIRS phantom produces doses within ±5% of the CT-based plan while CATPhan 504 produces a variation of ±14% of the CT-based plan. [ABSTRACT FROM AUTHOR]