Back to Search
Start Over
Characterization of Al2O3 sintered pellets for dosimetric applications in radiotherapy.
- Source :
-
Journal of radiological protection : official journal of the Society for Radiological Protection [J Radiol Prot] 1999 Mar; Vol. 19 (1), pp. 51-5. - Publication Year :
- 1999
-
Abstract
- Al2O3 sintered pellets were tested in relation to their thermoluminescent properties, to investigate the possibility of their use for dosimetry in the radiotherapy cases of 60Co and x-radiation. The material is inexpensive and has a glow peak at about 280 degrees C, a linear dose response between 0.1 and 100 Gy, suitable reproducibility and a low detection limit. The thermoluminescent response can easily be corrected for fading and the energy dependence, when necessary.
Details
- Language :
- English
- ISSN :
- 0952-4746
- Volume :
- 19
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of radiological protection : official journal of the Society for Radiological Protection
- Publication Type :
- Academic Journal
- Accession number :
- 10321696
- Full Text :
- https://doi.org/10.1088/0952-4746/19/1/006