Back to Search Start Over

[The application of the endotoxin test for globulin and other blood products].

Authors :
Fujiwara H
Kira K
Asakawa S
Naito S
Tanaka S
Aizawa M
Tuchiya M
Takaoka A
Yamamoto A
Ochiai M
Horiuchi Y
Source :
Yakugaku zasshi : Journal of the Pharmaceutical Society of Japan [Yakugaku Zasshi] 1999 Jun; Vol. 119 (6), pp. 451-5.
Publication Year :
1999

Abstract

The application of the endotoxin test for globulin and other blood products were investigated by two different limulus amebocyte lysate (LAL) test methods, colorimetric and kinetic turbidimetric methods, using two endotoxin-specific reagents. By the dilution of the blood products in 40 times or more, spiked endotoxin in the products was recovered accurately showing neither inhibition nor enhancement. The definite difference was not shown between the results obtained by the two LAL test methods. According to the method of the endotoxin test described under General Tests of The Japanese Pharmacopeia (thirteenth edition), JPXIII, the maximum valid dilution (MVD) for these products will be calculated to be 40 or more, so it is capable to measure the endotoxin limit for each product. This study indicates that the endotoxin test is applicable to measure the endotoxin content in globulin and other blood products as an alternative method for the rabbit pyrogen test.

Details

Language :
Japanese
ISSN :
0031-6903
Volume :
119
Issue :
6
Database :
MEDLINE
Journal :
Yakugaku zasshi : Journal of the Pharmaceutical Society of Japan
Publication Type :
Academic Journal
Accession number :
10376005
Full Text :
https://doi.org/10.1248/yakushi1947.119.6_451