Volume measurement using tissue characterization of three-dimensional endoscopic ultrasonographic images.

Authors :
Yoshino J
Nakazawa S
Inui K
Wakabayashi T
Okushima K
Kobayashi T
Nakamura S
Watanabe S
Asakura N
Source :
Endoscopy [Endoscopy] 2000 Aug; Vol. 32 (8), pp. 624-9.
Publication Year :
2000

Abstract

Background and Study Aims: Although it is time-consuming to measure the volume of lesions using three-dimensional endoscopic ultrasonography (3D-EUS), the technique is suitable for tissue characterization, as it allows images of multiple areas to be obtained simultaneously in uniform conditions. The present study tested automatic volume measurement using tissue characterization based on 3D-EUS.<br />Materials and Methods: Nine polygonal sections of resected spleen (volume 0.66 +/- 0.19 cm3) were immersed in water, and 40 radial 3D-EUS images were obtained. For tissue characterization, the methods of co-occurrence matrix and gray-level difference and discriminant analysis were used. Each spleen section was also measured using 3D-EUS. The volume of tissue identified as spleen using tissue characterization and the volume calculated on the basis of the 3D-EUS images were both compared with the actual volume measured beforehand. Measurements using tissue characterization and 3D-EUS were carried out for every third image. In three clinical cases of cancer the volume of the lesion was measured using tissue characterization and 3D-EUS.<br />Results: The mean volume of the nine splenic sections estimated using tissue characterization was 1.2 +/- 0.41 cm3 (mean +/- SD), while the mean volume estimated with 3D-EUS imaging was 1.1 +/- 0.30 cm3 (mean +/- SD). The volumes measured using tissue characterization were on average 13% larger than those obtained with 3D-EUS. Linear regression analysis showed a high degree of correlation between the two sets of measurements (r=0.97, P<0.00005), and also showed a high correlation between the volumes obtained using tissue characterization and the actual volume (r=0.93, P<0.0005). However, the volumes calculated using 3D-EUS images were larger than the actual volume (61% on average), and the volumes estimated using tissue characterization were also greater than the actual volume. The overestimation reflected the fact that measurement was only carried out in every third 3D-EUS image. In the clinical cases, the mean value for "true" tumor tissue as determined on EUS imaging represented 73% of the volume interpreted as cancer using tissue characterization.<br />Conclusions: There was a good correlation between the volume measured with 3D-EUS and the volume obtained using tissue characterization. The tissue characterization volumes were only relatively slightly larger than the volumes measured using 3D-EUS, suggesting that there may be some promise for this application of tissue characterization.

Details

Language :
English
ISSN :
0013-726X
Volume :
32
Issue :
8
Database :
MEDLINE
Journal :
Endoscopy
Publication Type :
Academic Journal
Accession number :
10935791
Full Text :
https://doi.org/10.1055/s-2000-9015