- Clinical applications of cardiac assist systems still suffer from thromboembolic complications due to thrombus formation behind the valve's leaflets. Wherever the flow is stagnant such as in the sinus, a thrombus is likely to form. The valve design in this study avoids stagnation zones behind the leaflets by a purge flow during systole. This purge flow is separated from the valve's main flow by a flow divider directing a part of the main flow into the sinus behind the leaflet. The optimization was performed on a monoleaflet valve because of its simple geometry. Relevant geometric parameters were systematically varied. Thirty-two models were designed and numerically investigated. The models with the best results were preselected and investigated in a computer-controlled valve tester. The washout of a dye filled in the sinus was digitally recorded and quantified. The results show that a sinus purge flow minimizes stagnation areas in the sinus and thus may avoid thrombus formation.