Basiliximab interferes with the detection of soluble IL-2 receptor by the Immulite Immunoassay system.

Authors :
Warlé MC
Kwekkeboom J
Tilanus HW
Metselaar HJ
Source :
Journal of immunological methods [J Immunol Methods] 2003 Apr 01; Vol. 275 (1-2), pp. 133-6.
Publication Year :
2003

Abstract

The use of the Immulite Automated Immunoassay system is becoming an increasingly popular method for accurate and fast measurement of sIL-2R in serum. Here, we report that detection of sIL-2R using the Immulite Immunoassay system in serum samples of patients who are treated with basiliximab, a monoclonal antibody against the alpha-chain of the IL-2 receptor, may lead to false measurements.

Details

Language :
English
ISSN :
0022-1759
Volume :
275
Issue :
1-2
Database :
MEDLINE
Journal :
Journal of immunological methods
Publication Type :
Academic Journal
Accession number :
12667677
Full Text :
https://doi.org/10.1016/s0022-1759(03)00016-4