Radio-frequency tissue ablation of the liver: in vivo and ex vivo experiments with four different systems.

Authors :
Denys AL
De Baere T
Kuoch V
Dupas B
Chevallier P
Madoff DC
Schnyder P
Doenz F
Source :
European radiology [Eur Radiol] 2003 Oct; Vol. 13 (10), pp. 2346-52. Date of Electronic Publication: 2003 Aug 27.
Publication Year :
2003

Abstract

The aim of this study was to test the efficacy of four different radio-frequency ablation (RFA) systems in normal hepatic parenchyma in large animals. The RFA was applied to pig livers in vivo and to calf livers ex vivo using the Radionics cluster needle, RITA starburst XL needle, Radiotherapeutics Le Veen 4.0 needle, and the Berchtold 14-G saline-perfused 15-mm active-tip needle based on constructor specifications. The volume of tissue coagulation from RF was calculated from measurements of the vertical diameter (Dv) and transverse diameter (Dt). Lesion shape was characterized using the ratio between Dt/Dv. Radiotherapeutics and RITA produced in vivo lesion volume of 42+/-10, 39+/-4 cm3 with a reproducible spherical shape (Dt/Dv of 1.01+/-0.16 and 0.97+/-0.1, respectively). Radionics produced in vivo RF lesions volume of 29+/-11 cm3) with an ovoid shape (Dt/Dv 0.88+/-0.09). The RF lesions with the Berchtold device could not be assessed in vivo as 5 of 8 animals died during treatment. Ex vivo RF lesions had similar volumes with each system; however, the Radiotherapeutics device produced more reproducible shaped lesions than the other systems. In our experimental study, we found no difference between expandable needle systems in vivo. Cooled needles produced slightly smaller and ovoid shape in vivo lesions.

Details

Language :
English
ISSN :
0938-7994
Volume :
13
Issue :
10
Database :
MEDLINE
Journal :
European radiology
Publication Type :
Academic Journal
Accession number :
12942277
Full Text :
https://doi.org/10.1007/s00330-003-1970-0