Back to Search Start Over

A pilot study to evaluate the SMART BAG: a new pressure-responsive, gas-flow limiting bag-valve-mask device.

Authors :
Wagner-Berger HG
Wenzel V
Voelckel WG
Rheinberger K
Stadlbauer KH
Müller T
Augenstein S
von Goedecke A
Lindner KH
Keller C
Source :
Anesthesia and analgesia [Anesth Analg] 2003 Dec; Vol. 97 (6), pp. 1686-1689.
Publication Year :
2003

Abstract

Unlabelled: Reducing inspiratory flow rate and peak airway pressure may be important to minimize the risk of stomach inflation when ventilating an unprotected airway with positive pressure ventilation. In this study, we assessed the effects of a standard self-inflating bag compared with a new pressure-responsive, inspiratory gas flow-limiting device (SMART BAG) on respiratory mechanics in 60 adult patients undergoing routine induction of anesthesia. Respiratory variables were measured using a pulmonary monitor. The SMART BAG resulted in significantly decreased inspiratory flow rate and peak airway pressure while providing adequate tidal volume delivery.<br />Implications: The SMART BAG, a new pressure-responsive, peak inspiratory gas flow-limiting bag-valve mask device, limits inspiratory gas flow from up to 120 L/min in a standard self-inflating bag to approximately 40 L/min. It is designed for use by all levels of health care professionals and has been proven in a clinical pilot study to effectively ventilate patients in respiratory arrest.

Details

Language :
English
ISSN :
0003-2999
Volume :
97
Issue :
6
Database :
MEDLINE
Journal :
Anesthesia and analgesia
Publication Type :
Academic Journal
Accession number :
14633543
Full Text :