Back to Search Start Over

Microwave thermal imaging: initial in vivo experience with a single heating zone.

Authors :
Meaney PM
Fanning MW
Paulsen KD
Lit D
Pendergrass SA
Fang Q
Moodie KL
Source :
International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group [Int J Hyperthermia] 2003 Nov-Dec; Vol. 19 (6), pp. 617-41.
Publication Year :
2003

Abstract

The deployment of hyperthermia as a routine adjuvant to radiation or chemotherapy is limited largely by the inability to devise treatment plans which can be monitored through temperature distribution feedback during therapy. A non-invasive microwave tomographic thermal imaging system is currently being developed which has previously exhibited excellent correlation between the recovered electrical conductivity of a heated zone and its actual temperature change during phantom studies. To extend the validation of this approach in vivo, the imaging system has been re-configured for small animal experiments to operate within the bore of a CT scanner for anatomical and thermometry registration. A series of 5-7 day old pigs have been imaged during hyperthermia with a monopole antenna array submerged in a saline tank where a small plastic tube surgically inserted the length of the abdomen has been used to create a zone of heated saline at pre-selected temperatures. Tomographic microwave data over the frequency range of 300-1000 MHz of the pig abdomen in the plane perpendicular to the torso is collected at regular intervals after the tube saline temperatures have settled to the desired settings. Images are reconstructed over a range of operating frequencies. The tube location is clearly visible and the recovered saline conductivity varies linearly with the controlled temperature values. Difference images utilizing the baseline state prior to heating reinforces the linear relationship between temperature and imaged saline conductivity. Demonstration of in vivo temperature recovery and correlation with an independent monitoring device is an important milestone prior to clinical integration of this non-invasive imaging system with a thermal therapy device.

Details

Language :
English
ISSN :
0265-6736
Volume :
19
Issue :
6
Database :
MEDLINE
Journal :
International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group
Publication Type :
Academic Journal
Accession number :
14756452
Full Text :
https://doi.org/10.1080/0265673031000140822