High-speed analysis of residual solvents by flow-modulation gas chromatography.
- Source :
-
Journal of chromatography. A [J Chromatogr A] 2004 Feb 20; Vol. 1027 (1-2), pp. 75-82. - Publication Year :
- 2004
-
Abstract
- High-speed gas chromatographic (GC) separation of residual solvents in pharmaceutical preparations, using a flow-modulation technique, is described. These volatile compounds are separated on a series-coupled (tandem) column ensemble consisting of a polyethylene glycol column and a trifluoropropylmethyl/dimethylpolysiloxane column. This column ensemble is operated in stop-flow mode to enhance, or "tune", the separation. A valve between the junction point of the tandem column ensemble and a source of carrier gas at a pressure above the GC inlet pressure is opened for intervals of 2-8 s. This stops or slightly reverses the flow of carrier gas in the first column. Stop-flow pulses are used to increase the separation of target analytes that overlap in the total ensemble chromatogram, compared to non-stop-flow, or conventional, operation. All 36 target compounds, based on ICH Classes I and II residual solvent lists, are resolved in 12 min using the stop-flow technique and a single chromatographic analysis.
Details
- Language :
- English
- ISSN :
- 0021-9673
- Volume :
- 1027
- Issue :
- 1-2
- Database :
- MEDLINE
- Journal :
- Journal of chromatography. A
- Publication Type :
- Academic Journal
- Accession number :
- 14971486
- Full Text :
- https://doi.org/10.1016/j.chroma.2003.10.046