Back to Search
Start Over
Ultrasensitive potentiometric immunosensor based on SA and OCA techniques for immobilization of HBsAb with colloidal Au and polyvinyl butyral as matrixes.
- Source :
-
Langmuir : the ACS journal of surfaces and colloids [Langmuir] 2004 Aug 17; Vol. 20 (17), pp. 7240-5. - Publication Year :
- 2004
-
Abstract
- A novel potentiometric immunosensor for detection of hepatitis B surface antigen (HBsAg) has been developed by means of self-assembly (SA) and opposite-charged adsorption (OCA) techniques to immobilize hepatitis B surface antibody (HBsAb) on a platinum electrode. A cleaned platinum electrode was first pretreated in the presence of 10% HNO3 and 2.5% K2CrO4 solution and held at -1.5 V (vs SCE) for 1 min to make it negatively charged and then immersed in a mixing solution containing hepatitis B surface antibody, colloidal gold (Au), and polyvinyl butyral (PVB). Finally, HBsAb was successfully immobilized onto the surface of the negatively charged platinum electrode modified nanosized gold and PVB sol-gel matrixes. The modified procedure was characterized by electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). The immobilized hepatitis B surface antibody exhibited direct electrochemical behavior toward hepatitis B surface antigen (HBsAg). The performance and factors influencing the performance of the resulting immunosensor were studied in detail. More than 95.7% of the results of the human serum samples obtained by this method were in agreement with those obtained by enzyme-linked immunosorbent assays (ELISAs). The resulting immunosensor exhibited fast potentiometric response (<3 min) to HBsAg. The detection limit of the immunosensor was 2.3 ng.mL(-1), and the linear range was from 8 to 1280 ng.mL(-1). Moreover, the studied immunosensor exhibited high sensitivity, good reproducibility, and long-term stability (>6 months).
- Subjects :
- Adsorption
Biosensing Techniques methods
Electrodes
Humans
Hydrogen-Ion Concentration
Platinum chemistry
Potentiometry instrumentation
Potentiometry methods
Sensitivity and Specificity
Serum chemistry
Surface Properties
Time Factors
Biosensing Techniques instrumentation
Colloids chemistry
Gold chemistry
Hepatitis B Antibodies chemistry
Polyvinyls chemistry
Subjects
Details
- Language :
- English
- ISSN :
- 0743-7463
- Volume :
- 20
- Issue :
- 17
- Database :
- MEDLINE
- Journal :
- Langmuir : the ACS journal of surfaces and colloids
- Publication Type :
- Academic Journal
- Accession number :
- 15301511
- Full Text :
- https://doi.org/10.1021/la030428m