Reference threshold levels for an ER-3A insert earphone.

Authors :
Frank T
Vavrek MJ
Source :
Journal of the American Academy of Audiology [J Am Acad Audiol] 1992 Jan; Vol. 3 (1), pp. 51-9.
Publication Year :
1992

Abstract

The purpose of this study was to obtain normal hearing thresholds for an ER-3A insert earphone as a contribution for future standardization and for comparison with previous studies and interim ANSI and ISO insert earphone reference equivalent threshold sound pressure levels (RETSPLs). Hearing thresholds were obtained on each ear of 48 normal-hearing subjects from 125 to 8000 Hz using ER-3A and TDH-49P earphones referenced to a Bruel and Kjaer DB-0138 HA-2 and NBS-9A acoustic coupler, respectively. The mean unadjusted and adjusted ER-3A thresholds were in very good agreement with the mean thresholds averaged over eight other studies. Further, the mean adjusted ER-3A thresholds were in very good agreement with the interim ANSI RETSPLs while the mean unadjusted ER-3A thresholds were slightly higher than the ISO RETSPLs. Also, mean ER-3A thresholds averaged over a nine study data base were in closer agreement with the interim ANSI than the ISO RETSPLS. Overall, the interim ANSI insert earphone RETSPLs were recommended for clinical use.

Details

Language :
English
ISSN :
1050-0545
Volume :
3
Issue :
1
Database :
MEDLINE
Journal :
Journal of the American Academy of Audiology
Publication Type :
Academic Journal
Accession number :