Back to Search Start Over

Evaluation of automated sex hormone binding globulin immunoassays.

Authors :
Jin M
Wener MH
Bankson DD
Source :
Clinical biochemistry [Clin Biochem] 2006 Jan; Vol. 39 (1), pp. 91-4. Date of Electronic Publication: 2005 Dec 02.
Publication Year :
2006

Abstract

Background: Sex hormone binding globulin (SHBG) is an important regulator of testosterone and estradiol.<br />Study Design: We validated the Diagnostic Products Corporation (DPC) and Roche Diagnostic SHBG immunoassays on the DPC Immulite 2000 and Roche Modular E170 analyzers.<br />Results: The coefficient of variation for SHBG kits from both manufacturers was in the range of 3.9-7.7% (between-run) and 0.95-5.0% (within-run), free of interference from hemoglobin, bilirubin, lipid, and rheumatoid factor, and linear up to at least 170 nM SHBG. The results of the two methods, however, were biased by up to 29% depending on the SHBG concentration.<br />Conclusion: The SHBG assays perform well but standardization is needed.

Details

Language :
English
ISSN :
0009-9120
Volume :
39
Issue :
1
Database :
MEDLINE
Journal :
Clinical biochemistry
Publication Type :
Academic Journal
Accession number :
16325792
Full Text :
https://doi.org/10.1016/j.clinbiochem.2005.10.012