Back to Search Start Over

Activity measurement of a 176Lu sample using coincidence peaks and Monte Carlo simulations.

Authors :
Jutier C
Le Petit G
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2006 Oct-Nov; Vol. 64 (10-11), pp. 1292-6. Date of Electronic Publication: 2006 Apr 17.
Publication Year :
2006

Abstract

We present the results of the activity measurement of a sample of the rare isotope 176Lu, obtained by gamma-ray spectrometry through the use of 22 peaks in an energy spectrum acquired with a 300 cc well-type HPGe detector located in the Modane underground facility below the French Alps. Of the above peaks, 17 are coincidence peaks between two or three gamma-rays or X-rays. True coincidence summing corrections were made using our Coincal code. Monte Carlo simulations were carried out for calibration purposes.