Back to Search Start Over

An animal model in sheep for biocompatibility testing of biomaterials in cancellous bones.

Authors :
Nuss KM
Auer JA
Boos A
von Rechenberg B
Source :
BMC musculoskeletal disorders [BMC Musculoskelet Disord] 2006 Aug 15; Vol. 7, pp. 67. Date of Electronic Publication: 2006 Aug 15.
Publication Year :
2006

Abstract

Background: The past years have seen the development of many synthetic bone replacements. To test their biocompatibility and ability for osseointegration, osseoinduction and -conduction requires their placement within bone preferably in an animal experiment of a higher species.<br />Methods: A suitable experimental animal model in sheep with drill holes of 8 mm diameter and 13 mm depth within the proximal and distal humerus and femur for testing biocompatibility issues is introduced.<br />Results: This present sheep model allows the placing of up to 8 different test materials within one animal and because of the standardization of the bone defect, routine evaluation by means of histomorphometry is easily conducted. This method was used successfully in 66 White Alpine Sheep. When the drill holes were correctly placed no complications such as spontaneous fractures were encountered.<br />Conclusion: This experimental animal model serves an excellent basis for testing the biocompatibility of novel biomaterials to be used as bone replacement or new bone formation enhancing materials.

Details

Language :
English
ISSN :
1471-2474
Volume :
7
Database :
MEDLINE
Journal :
BMC musculoskeletal disorders
Publication Type :
Academic Journal
Accession number :
16911787
Full Text :
https://doi.org/10.1186/1471-2474-7-67