Back to Search Start Over

Two-step hard X-ray focusing combining Fresnel zone plate and single-bounce ellipsoidal capillary.

Authors :
Snigirev A
Bjeoumikhov A
Erko A
Snigireva I
Grigoriev M
Yunkin V
Erko M
Bjeoumikhova S
Source :
Journal of synchrotron radiation [J Synchrotron Radiat] 2007 Jul; Vol. 14 (Pt 4), pp. 326-30. Date of Electronic Publication: 2007 Jun 14.
Publication Year :
2007

Abstract

A two-step focusing set-up combining a Fresnel zone plate with an ellipsoidal capillary is presented. It is shown that, in addition to the anticipated gain in flux, the employment of the prefocusing micro-optic makes optimal use of the elliptical shape of the capillary by almost eliminating aberrations. A small cross section of the prefocused beam allows a tiny fraction of the capillary surface to be selected, thus reducing the influence of slope errors. An X-ray beam with a 15 keV energy was focused down to a spot size as small as 250 nm, demonstrating the best value that has been achieved up to now for single-bounce capillaries. The use of an ellipsoidal capillary as a micromirror under off-axis illumination by microfocusing optics may open up new opportunities in nanofocusing developments.

Details

Language :
English
ISSN :
0909-0495
Volume :
14
Issue :
Pt 4
Database :
MEDLINE
Journal :
Journal of synchrotron radiation
Publication Type :
Academic Journal
Accession number :
17587657
Full Text :
https://doi.org/10.1107/S0909049507025174