Back to Search Start Over

A radiofrequency-assisted device for bloodless rapid transection of the liver: a comparative study in a pig liver model.

Authors :
Burdío F
Navarro A
Berjano E
Sousa R
Burdío JM
Güemes A
Subiró J
Gonzalez A
Cruz I
Castiella T
Tejero E
Lozano R
Grande L
de Gregorio MA
Source :
European journal of surgical oncology : the journal of the European Society of Surgical Oncology and the British Association of Surgical Oncology [Eur J Surg Oncol] 2008 May; Vol. 34 (5), pp. 599-605. Date of Electronic Publication: 2007 Jul 05.
Publication Year :
2008

Abstract

Background: Efficient and safe liver parenchymal transection is dependent on the ability to address both parenchymal division and hemostasis simultaneously. In this article we describe and compare with a saline-linked instrument a new radiofrequency (RF)-assisted device specifically designed for tissue thermocoagulation and division of the liver used on an in vivo pig liver model.<br />Methods: In total, 20 partial hepatectomies were performed on pigs through laparotomy. Two groups were studied: group A (n=8) with hepatectomy performed using only the proposed RF-assisted device and group B (n=8) with hepatectomy performed using only a saline-linked device. Main outcome measures were: transection time, blood loss during transection, transection area, transection speed and blood loss per transection area. Secondary measures were: risk of biliary leakage, tissue coagulation depth and the need for hemostatic stitches. Tissue viability was evaluated in selected samples by staining of tissue NADH.<br />Results: In group A both blood loss and blood loss per transection area were lower (p=0.001) than in group B (70+/-74 ml and 2+/-2 ml/cm(2) vs. 527+/-273 ml and 13+/-6 ml/cm(2), for groups A and B, respectively). An increase in mean transection speed when using the proposed device over the saline-linked device group was also demonstrated (3+/-0 and 2+/-1cm(2)/min for group A and B, respectively) (p=0.002). Tissue coagulation depth was greater (p=0.005) in group A than in group B (6+/-2 mm and 3+/-1 mm, for groups A and B, respectively). Neither macroscopic nor microscopic differences were encountered in transection surfaces between both groups.<br />Conclusions: The proposed RF-assisted device was shown to address parenchymal division and hemostasis simultaneously, with less blood loss and faster transection time than saline-linked technology in this experimental model.

Details

Language :
English
ISSN :
1532-2157
Volume :
34
Issue :
5
Database :
MEDLINE
Journal :
European journal of surgical oncology : the journal of the European Society of Surgical Oncology and the British Association of Surgical Oncology
Publication Type :
Academic Journal
Accession number :
17614248
Full Text :
https://doi.org/10.1016/j.ejso.2007.05.008