Back to Search
Start Over
Automated capture and on-column detection of biotinylated DNA on a disposable solid support.
- Source :
-
The Analyst [Analyst] 2007 Aug; Vol. 132 (8), pp. 818-22. Date of Electronic Publication: 2007 Jul 02. - Publication Year :
- 2007
-
Abstract
- This work comprises the development of a technique for the capture of single-stranded DNA on a solid support combined with in situ quantification. The capture is based on the strong and selective interaction between biotinylated DNA and streptavidin-coated agarose beads. Sequential Injection in the lab-on-valve format allows for automated manipulation of all components including the building and disposal of bead columns. Detection was accomplished using the OliGreen fluorescent dye and optimization of the assay achieved a limit of detection of 111 pg ssDNA, with a total assay time of roughly 2.5 min per sample.
- Subjects :
- Autoanalysis instrumentation
Biotin
Calibration
Flow Injection Analysis instrumentation
Flow Injection Analysis methods
Fluorescent Dyes
Humans
Microfluidic Analytical Techniques instrumentation
Microfluidic Analytical Techniques methods
Nanoparticles
Nucleic Acid Denaturation
Autoanalysis methods
DNA, Single-Stranded analysis
Subjects
Details
- Language :
- English
- ISSN :
- 0003-2654
- Volume :
- 132
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- The Analyst
- Publication Type :
- Academic Journal
- Accession number :
- 17646882
- Full Text :