Back to Search Start Over

A local region of interest image reconstruction via filtered backprojection for fan-beam differential phase-contrast computed tomography.

Authors :
Qi Z
Chen GH
Source :
Physics in medicine and biology [Phys Med Biol] 2007 Sep 21; Vol. 52 (18), pp. N417-23. Date of Electronic Publication: 2007 Sep 04.
Publication Year :
2007

Abstract

Recently, x-ray differential phase contrast computed tomography (DPC-CT) has been experimentally implemented using a conventional source combined with several gratings. Images were reconstructed using a parallel-beam reconstruction formula. However, parallel-beam reconstruction formulae are not directly applicable for a large image object where the parallel-beam approximation fails. In this note, we present a new image reconstruction formula for fan-beam DPC-CT. There are two major features in this algorithm: (1) it enables the reconstruction of a local region of interest (ROI) using data acquired from an angular interval shorter than 180 degrees + fan angle and (2) it still preserves the filtered backprojection structure. Numerical simulations have been conducted to validate the image reconstruction algorithm.

Details

Language :
English
ISSN :
0031-9155
Volume :
52
Issue :
18
Database :
MEDLINE
Journal :
Physics in medicine and biology
Publication Type :
Academic Journal
Accession number :
17804875
Full Text :
https://doi.org/10.1088/0031-9155/52/18/N03