Back to Search Start Over

Closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study.

Authors :
Ngan Kee WD
Tam YH
Khaw KS
Ng FF
Critchley LA
Karmakar MK
Source :
Anaesthesia [Anaesthesia] 2007 Dec; Vol. 62 (12), pp. 1251-6.
Publication Year :
2007

Abstract

We describe the novel use of a closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure in 53 patients having spinal anaesthesia for elective caesarean section. A simple on-off algorithm was used that activated an intravenous phenylephrine infusion at 100 microg.min(-1) when systolic blood pressure was less than or equal to baseline and stopped the infusion when systolic blood pressure exceeded baseline. Up to uterine incision, 94.6% of all systolic blood pressure measurements were within the range (baseline +/- 20%). Seven patients (13.2%) had one or more episodes of hypotension (systolic blood pressure < 80% of baseline) and 23 patients (37.7%) had one or more episodes of hypertension (systolic blood pressure > 120% of baseline). No patient had nausea or vomiting and in no case was umbilical arterial blood pH < 7.2. Calculated system performance parameters were comparable with those of previously published closed-loop systems and provide a reference for the potential development and comparison of more advanced algorithms.

Details

Language :
English
ISSN :
1365-2044
Volume :
62
Issue :
12
Database :
MEDLINE
Journal :
Anaesthesia
Publication Type :
Academic Journal
Accession number :
17991262
Full Text :