Radiofrequency hepatic ablation with internally cooled electrodes and hybrid applicators with distant saline infusion using an in vivo porcine model.

Authors :
Burdío F
Navarro A
Berjano EJ
Burdío JM
Gonzalez A
Güemes A
Sousa R
Rufas M
Cruz I
Castiella T
Lozano R
Lequerica JL
Grande L
Source :
European journal of surgical oncology : the journal of the European Society of Surgical Oncology and the British Association of Surgical Oncology [Eur J Surg Oncol] 2008 Jul; Vol. 34 (7), pp. 822-30. Date of Electronic Publication: 2007 Nov 26.
Publication Year :
2008

Abstract

Aims: Radiofrequency ablation (RFA) of tumors by means of internally cooled (ICE) or multitined expandable electrodes combined with infusion of saline into the tissue may improve results. Our aim was to determine the efficacy of a previously optimized hybrid ICE system (ICE combined with infusion of saline into the tissue at a distance of 2mm) in comparison with a conventional ICE cluster electrode in porcine liver in vivo.<br />Methods: A total of 32 RFA were performed on a total of 10 farm pigs using two RFA systems: Group A (n=16): Cluster electrode. Group B (n=16): Hybrid system (with continuous infusion of 100ml/h of 20% NaCl at 2mm distance from the electrode shaft by an independent isolated needle). Livers were removed for macroscopic and histological assessment after the procedure. Coagulation volume, coagulation diameters, coefficient of variability (CV) of coagulation volume, sphericity ratio (SR), deposited power (DP), deposited energy (DE), deposited energy per coagulation volume (DEV) and rise of animal temperature during the procedure were compared and correlated among groups. Additionally, linear regression analysis was modeled to study the relationship between deposited energy and either coagulation volume and rise of animal temperature during the procedure in both groups.<br />Results: Both coagulation volume and short diameter of coagulation were significantly greater (p<0.05) in group B compared to group A (22.7+/-11.0 cm(3) and 3.1+/-0.7 cm vs. 13.5+/-7.7 cm(3) and 2.5+/-0.5 cm, respectively). A similar CV and SR was observed among groups (57.1% and 1.4+/-0.3 vs. 48.6% and 1.3+/-0.2 for groups B and A, respectively). In group B, DE and DP were more than double group A, but DEV was nearly twice as high (9782 J/cm(3) vs. 5342 J/cm(3), for groups B and A, respectively). No significant relationship between DE and coagulation volume was encountered.<br />Conclusion: Efficacy of a single ICE may be improved with continuous infusion of saline at around 2 mm from the electrode shaft. Coagulation volume obtained with this improved system may be even greater than that obtained with a cluster electrode.

Details

Language :
English
ISSN :
1532-2157
Volume :
34
Issue :
7
Database :
MEDLINE
Journal :
European journal of surgical oncology : the journal of the European Society of Surgical Oncology and the British Association of Surgical Oncology
Publication Type :
Academic Journal
Accession number :
18042500
Full Text :
https://doi.org/10.1016/j.ejso.2007.09.029