[Evaluation of the Abbott Cell-Dyn Sapphire hematology analyzer].

Authors :
Park Y
Song J
Song S
Song KS
Ahn MS
Yang MS
Kim I
Choi JR
Source :
The Korean journal of laboratory medicine [Korean J Lab Med] 2007 Jun; Vol. 27 (3), pp. 162-8.
Publication Year :
2007

Abstract

Background: The performance of Cell-Dyn Sapphire (Abbott Diagnostic, USA) was compared to the Bayer Advia 2120 (Bayer Diagnostics, USA), Sysmex XE-2100 (Sysmex Corporation, Japan), and reference microscopy.<br />Methods: Three hundred samples for routine CBC and WBC differentials were randomly chosen for a comparison analysis. The Cell-Dyn Sapphire system was evaluated according to the linearity, imprecision, inter-instrument correlations, and white blood cell differential.<br />Results: The CBC parameters (WBC, RBC, hemoglobin and platelet) showed a significant linearity with correlation coefficients greater than 0.99 (P<0.0001). Coefficients of variation (CV) for within-run and differential count of WBC were less than 5% except for Total CV for monocytes, eosinophils, and basophils and within-run CV for low valued eosinophils. The correlation coefficients with manual count were lower in monocytes, eosinophils, and basophils than in neutrophils and lymphocytes. The correlation with other hematology anlayzers was significant exclusive of basophils.<br />Conclusions: These results demonstrate that the Cell-Dyn Sapphire has a good linearity, an acceptable reproducibility, a minimal carryover, and a comparable performance with the sysmex XE-2100 and Advia 2120.

Details

Language :
Korean
ISSN :
1598-6535
Volume :
27
Issue :
3
Database :
MEDLINE
Journal :
The Korean journal of laboratory medicine
Publication Type :
Academic Journal
Accession number :
18094570
Full Text :
https://doi.org/10.3343/kjlm.2007.27.3.162