Back to Search
Start Over
The use of photostimulable phosphor systems for periodic quality assurance in radiotherapy.
- Source :
-
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) [Phys Med] 2008 Mar; Vol. 24 (1), pp. 34-7. - Publication Year :
- 2008
-
Abstract
- The fusion of radiological and optical images can be achieved through charging a photostimulable phosphor plate (PSP) with an exposure to a field of X- or gamma-rays, followed by exposure to an optical image which discharges the plate in relation to the amount of incident light. According to this PSP characteristic, we developed a simple method for periodic quality assurance (QA) of light/radiation field coincidence, distance indicator, field size indicators, crosshair centering, coincidence of radiation and mechanical isocenter for linear accelerators. The geometrical accuracy of radiological units can be subjected to the same QA method. Further, the source position accuracy for an HDR remote afterloader can be checked by taking an autoradiography of the radioactive source and simultaneously an optical image of a reference geometrical system.
- Subjects :
- Equipment Design
Equipment Failure Analysis
Light
Radiometry methods
Radiotherapy methods
Reproducibility of Results
Sensitivity and Specificity
Phosphorus radiation effects
Quality Assurance, Health Care methods
Radiometry instrumentation
Radiotherapy instrumentation
X-Ray Intensifying Screens
Subjects
Details
- Language :
- English
- ISSN :
- 1120-1797
- Volume :
- 24
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
- Publication Type :
- Academic Journal
- Accession number :
- 18178116
- Full Text :
- https://doi.org/10.1016/j.ejmp.2007.11.003