The reliability of computed tomography (CT) values and dimensional measurements of the oropharyngeal region using cone beam CT: comparison with multidetector CT.
- Source :
-
Dento maxillo facial radiology [Dentomaxillofac Radiol] 2008 Jul; Vol. 37 (5), pp. 245-51. - Publication Year :
- 2008
-
Abstract
- Objectives: To evaluate the reliability of cone beam CT (CBCT) values and dimensional measurements of oropharyngeal air spaces as compared with those of multidetector row CT (MDCT).<br />Methods: A soft tissue equivalent phantom with different sized holes was used. The phantom was scanned using CBCT and MDCT. In addition, a volunteer was examined using both methods for clinical comparison. The CT data were retrieved to a personal computer and treated using image processing software for three-dimensional rendering and dimensional measurements. The CT values of air, water and soft tissues were measured experimentally and the CT values of air, fat and muscle were measured in a clinical case. The diameter of each hole drilled in the phantom was measured on CBCT and MDCT images using the software tool and digimatic callipers.<br />Results: The CBCT values were quite different from the Hounsfield units obtained with a typical MDCT system. The standard deviations were almost ten times larger with CBCT. In vivo assessment showed that the CBCT values for fat had a wide range that partially overlapped the values for muscle. The measurement of distances greater than 4 mm was consistent for all methods. The difference for holes was less than 0.2 mm.<br />Conclusions: The phantom study showed that CBCT provides limited quantitative CT values for each pixel on sliced images for differentiating air, water and soft tissues. However, the measurement of air spaces with CBCT was quite accurate.
- Subjects :
- Adipose Tissue diagnostic imaging
Air
Cephalometry standards
Epoxy Resins
Humans
Imaging, Three-Dimensional methods
Imaging, Three-Dimensional standards
Masseter Muscle diagnostic imaging
Microcomputers
Neck Muscles diagnostic imaging
Phantoms, Imaging
Pterygoid Muscles diagnostic imaging
Radiographic Image Enhancement standards
Reproducibility of Results
Submandibular Gland diagnostic imaging
Tomography, X-Ray Computed methods
Water
Cone-Beam Computed Tomography standards
Image Processing, Computer-Assisted standards
Oropharynx diagnostic imaging
Tomography, X-Ray Computed standards
Subjects
Details
- Language :
- English
- ISSN :
- 0250-832X
- Volume :
- 37
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Dento maxillo facial radiology
- Publication Type :
- Academic Journal
- Accession number :
- 18606745
- Full Text :
- https://doi.org/10.1259/dmfr/45926904