Back to Search Start Over

A Bremsstrahlung spectrometer using k-edge and differential filters with image plate dosimeters.

Authors :
Chen CD
King JA
Key MH
Akli KU
Beg FN
Chen H
Freeman RR
Link A
Mackinnon AJ
MacPhee AG
Patel PK
Porkolab M
Stephens RB
Van Woerkom LD
Source :
The Review of scientific instruments [Rev Sci Instrum] 2008 Oct; Vol. 79 (10), pp. 10E305.
Publication Year :
2008

Abstract

A Bremsstrahlung spectrometer using k-edge and differential filtering has been used with image plate dosimeters to measure the x-ray fluence from short-pulse laser/target interactions. An electron spectrometer in front of the Bremsstrahlung spectrometer deflects electrons from the x-ray line of sight and simultaneously measures the electron spectrum. The response functions were modeled with the Monte Carlo code INTEGRATED TIGER SERIES 3.0 and the dosimeters calibrated with radioactive sources. An electron distribution with a slope temperature of 1.3 MeV is inferred from the Bremsstrahlung spectra.

Details

Language :
English
ISSN :
1089-7623
Volume :
79
Issue :
10
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
19044467
Full Text :
https://doi.org/10.1063/1.2964231