Back to Search Start Over

A hybrid method to determine efficiency curve of HPGe detectors.

Authors :
Ozben CS
Emirhan EM
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2009 Jun; Vol. 67 (6), pp. 1110-3. Date of Electronic Publication: 2009 Feb 10.
Publication Year :
2009

Abstract

A method was developed for determining the efficiency calibration curve of HPGe detectors between 26 and 1770 keV. A simple simulation based on the attenuation of gamma-rays in pure germanium was used together with efficiency calibrations performed for three detector-source distances: 5, 12 and 30 cm. The simple simulation model required an energy dependent term with three parameters and these parameters were determined from the fit of the simulated data to the measured efficiency of the detector. The efficiency response of the detector determined this way was then tested with other certified sources to verify the validity of the method.

Details

Language :
English
ISSN :
1872-9800
Volume :
67
Issue :
6
Database :
MEDLINE
Journal :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
Publication Type :
Academic Journal
Accession number :
19297177
Full Text :
https://doi.org/10.1016/j.apradiso.2009.01.085