Back to Search Start Over

Evaluation of a new portable glucose meter designed for the use in cats.

Authors :
Zini E
Moretti S
Tschuor F
Reusch CE
Source :
Schweizer Archiv fur Tierheilkunde [Schweiz Arch Tierheilkd] 2009 Sep; Vol. 151 (9), pp. 448-51.
Publication Year :
2009

Abstract

Portable blood glucose meters (PBGMs) are useful in the management of diabetes mellitus in cats. In the present study we compared the performance of two PBGMs: the AlphaTRAK (Abbott Animal Health, Maidenhead, England) specifically developed for dogs and cats, and the Ascensia ELITE (Bayer HealthCare, Zurich, Switzerland) developed for humans. Quality parameters, including precision and accuracy, were better for the AlphaTRAK meter compared to Ascensia ELITE. While the AlphaTRAK meter results did not differ from the reference method, results from the Ascensia ELITE were significantly (P<0.001) lower. The superior performance of the AlphaTRAK meter supports its use to monitor blood glucose levels in cats.

Details

Language :
English
ISSN :
0036-7281
Volume :
151
Issue :
9
Database :
MEDLINE
Journal :
Schweizer Archiv fur Tierheilkunde
Publication Type :
Academic Journal
Accession number :
19722134
Full Text :
https://doi.org/10.1024/0036-7281.151.9.448