Back to Search Start Over

Capacitive on-line hematocrit sensor design based on impedance spectroscopy for use in hemodialysis machines.

Authors :
Trebbels D
Hradetzky D
Zengerle R
Source :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2009; Vol. 2009, pp. 1208-11.
Publication Year :
2009

Abstract

This paper presents a new design for an on-line and in-line hematocrit (HCT) sensor. Special feature of the sensor is the capability to measure the hematocrit of a blood sample inside standard plastic tubing widely used in medical equipment. No blood sample has to be extracted out of existing extracorporeal blood circulation systems such as hemodialysis machines or heart-lung machines. The sensor principle is based on electrical impedance spectroscopy. Dielectric properties of the blood and the plastic tubing are measured at various frequencies. In order to optimize the sensitivity, a unique electrode configuration is developed and optimized by Finite Element Simulation. The new electrode design optimizes the overall sensitivity of the sensor towards a change in dielectric properties of the blood caused by the HCT value and therefore decreases the sensitivity to side effects caused by temperature drift and component tolerances. As a result of the optimized overall sensor performance the complexity of a sensor readout circuitry can be reduced to a minimum which leads to an unmatched price-performance ratio for a complete measurement system.

Details

Language :
English
ISSN :
2375-7477
Volume :
2009
Database :
MEDLINE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Publication Type :
Academic Journal
Accession number :
19963491
Full Text :
https://doi.org/10.1109/IEMBS.2009.5332420