The longevity of arteriovenous graft for hemodialysis patients--externally supported or nonsupported.
- Source :
-
Clinical journal of the American Society of Nephrology : CJASN [Clin J Am Soc Nephrol] 2010 Jun; Vol. 5 (6), pp. 1029-35. Date of Electronic Publication: 2010 Apr 08. - Publication Year :
- 2010
-
Abstract
- Background and Objectives: Because thrombosis and stenosis are the most frequent causes of arteriovenous graft (AVG) failure, several externally supported grafts were designed to enhance incompressibility and kinking resistance to improve the patency. However, consensus on advantage of these modifications having still not been reached might relate to the previous limited patient numbers and small studies. This study analyzed the longevity of externally supported and nonsupported AVG on the basis of the nationwide database to understand whether the externally supported design could improve the duration of patency of AVG.<br />Design, Setting, Participants, & Measurements: Adult incident patients (n = 4402) were retrospectively enrolled who had began hemodialysis between January 1, 2002 and December 31, 2005. All incident patients with AVG, before or after beginning regular hemodialysis, between January 1, 2000 and December 31, 2006 were identified. The Cox proportional hazard model was used to compare the longevity of externally supported and nonsupported AVG by controlling other covariates.<br />Results: There were 990 patients with externally supported AVG and 3412 patients with nonsupported AVG. The patency rates of externally supported AVG were 49.4%, 31.6%, and 20.2% at 1, 2 and 3 years, respectively; those of nonsupported AVG were 31.9%, 17.4%, and 10.8%. The survival of nonsupported AVG was significantly inferior to that of externally supported AVG.<br />Conclusions: Externally supported AVG are superior to nonsupported AVG for long-term patency. The incompressibility and kinking resistance of ringed grafts may shed light on a direction for the future development of dialysis graft design.
- Subjects :
- Adolescent
Adult
Aged
Blood Vessel Prosthesis
Blood Vessel Prosthesis Implantation adverse effects
Blood Vessel Prosthesis Implantation instrumentation
Chi-Square Distribution
Female
Graft Occlusion, Vascular etiology
Graft Occlusion, Vascular physiopathology
Humans
Male
Middle Aged
Proportional Hazards Models
Prosthesis Design
Retrospective Studies
Risk Assessment
Risk Factors
Taiwan
Time Factors
Treatment Outcome
Young Adult
Arteriovenous Shunt, Surgical adverse effects
Arteriovenous Shunt, Surgical instrumentation
Graft Occlusion, Vascular prevention & control
Renal Dialysis
Upper Extremity blood supply
Vascular Patency
Subjects
Details
- Language :
- English
- ISSN :
- 1555-905X
- Volume :
- 5
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Clinical journal of the American Society of Nephrology : CJASN
- Publication Type :
- Academic Journal
- Accession number :
- 20378642
- Full Text :
- https://doi.org/10.2215/CJN.08181109