Back to Search
Start Over
Optimizing clinical performance and geometrical robustness of a new electrode device for intracranial tumor electroporation.
- Source :
-
Bioelectrochemistry (Amsterdam, Netherlands) [Bioelectrochemistry] 2011 Apr; Vol. 81 (1), pp. 10-6. Date of Electronic Publication: 2011 Jan 06. - Publication Year :
- 2011
-
Abstract
- Current technology has limited applicability for electroporation based treatment of deep-seated tumors, and is in general, not optimized in terms of compliance with clinically relevant parameters. Here we present a novel electrode device developed for electrotransfer of antineoplastic drugs and genes to intracranial tumors in humans, and demonstrate a method to optimize the design (i.e. geometry) of the electrode device prototype to improve both clinical performance and geometrical tolerance (robustness). We have employed a semiempirical objective function based on constraints similar to those used in radiation oncology. The results show that small geometrical changes may yield a significant improvement. For example, a 2 mm displacement of 6 electrodes yields 14% better compliance with the clinical parameters, compared to the prototype, and additionally makes the electrode device less sensitive to random geometrical deviations. The method is readily applicable to other electrode configurations.<br /> (Copyright © 2011 Elsevier B.V. All rights reserved.)
- Subjects :
- Algorithms
Antineoplastic Agents administration & dosage
Brain Neoplasms surgery
Craniotomy
Drug Delivery Systems instrumentation
Electrochemotherapy instrumentation
Electrodes
Genetic Therapy methods
Tumor Burden
Brain Neoplasms drug therapy
Drug Delivery Systems methods
Electrochemotherapy methods
Subjects
Details
- Language :
- English
- ISSN :
- 1878-562X
- Volume :
- 81
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Bioelectrochemistry (Amsterdam, Netherlands)
- Publication Type :
- Academic Journal
- Accession number :
- 21256816
- Full Text :
- https://doi.org/10.1016/j.bioelechem.2010.12.002