Comparisons of multiple automated anatomy-based image-guidance methods for patient setup before head/neck external beam radiotherapy.
- Source :
-
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2010 Oct 13; Vol. 12 (1), pp. 3337. Date of Electronic Publication: 2010 Oct 13. - Publication Year :
- 2010
-
Abstract
- The purpose was to assess the variability in automated translational head/neck setup corrections computed from several different imaging modalities and rigid registration methods using patient anatomy. Shifts were calculated using three commercial and one in-house automated rigid registration methods for nine head/neck patients who were imaged with three different image-guidance systems. The mean difference between the daily isocenter shifts determined by the four methods ranged from 2.8 to 12.5 mm for all of the test cases.These differences are much greater than the variability observed for a rigid imaging phantom. Image-guided setup procedures have an uncertainty that depends on the imaging modality, the registration algorithm, the image resolution and the image content. In the absence of an absolute ground truth, the variation in the shifts calculated by several different methods provides a useful estimate of that uncertainty.