Back to Search Start Over

Reliability and validity of 4D rasterstereography under dynamic conditions.

Authors :
Betsch M
Wild M
Jungbluth P
Hakimi M
Windolf J
Haex B
Horstmann T
Rapp W
Source :
Computers in biology and medicine [Comput Biol Med] 2011 Jun; Vol. 41 (6), pp. 308-12. Date of Electronic Publication: 2011 Apr 12.
Publication Year :
2011

Abstract

Purpose of this study was to evaluate the reliability and validity of 4D rasterstereography under dynamic conditions. Therefore simulated anatomical fixed points on a wooden plate were measured during different movements. Seven different motion patterns in all three angles of space were evaluated. The simulated parameter trunk length was measured with an accuracy of 3.58 mm (SD±3.29 mm) and the dimple distance was detected with an accuracy of 0.88 mm (SD±1.04 mm). With this rasterstereographic prototype it is possible to examine dynamically the spinal posture with adequate accuracy.<br /> (Copyright © 2011 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1879-0534
Volume :
41
Issue :
6
Database :
MEDLINE
Journal :
Computers in biology and medicine
Publication Type :
Academic Journal
Accession number :
21489425
Full Text :
https://doi.org/10.1016/j.compbiomed.2011.03.008