Back to Search Start Over

Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass.

Authors :
Wong JH
Hardcastle N
Tomé WA
Bayliss A
Tolakanahalli R
Lerch ML
Petasecca M
Carolan M
Metcalfe P
Rosenfeld AB
Source :
Medical physics [Med Phys] 2011 Apr; Vol. 38 (4), pp. 2256-64.
Publication Year :
2011

Abstract

Purpose: Helical tomotherapy is a complex delivery technique, integrating CT image guidance and intensity modulated radiotherapy in a single system. The integration of the CT detector ring on the gantry not only allows patient position verification but is also often used to perform various QA procedures. This convenience lacks the rigor of a machine-independent QA process.<br />Methods: In this article, a Si strip detector, known as the Dose Magnifying Glass (DMG), was used to perform machine-independent QA measurements of the multileaf collimator alignment, leaf open time threshold, and leaf fluence output factor (LFOF).<br />Results: The DMG measurements showed good agreements with EDR2 film for the MLC alignment test while the CT detector agrees well with DMG measurements for leaf open time threshold and LFOF measurements. The leaf open time threshold was found to be approximately 20 ms. The LFOF measured with the DMG agreed within error with the CT detector measured LFOF.<br />Conclusions: The DMG with its 0.2 mm spatial resolution coupled to TERA ASIC allowed real-time high temporal resolution measurements of the tomotherapy leaf movement. In conclusion, DMG was shown to be a suitable tool for machine-independent QA of a tomotherapy unit.

Details

Language :
English
ISSN :
0094-2405
Volume :
38
Issue :
4
Database :
MEDLINE
Journal :
Medical physics
Publication Type :
Academic Journal
Accession number :
21626960
Full Text :
https://doi.org/10.1118/1.3566067