[Development of software for estimating exposure dose and radiation exposure region in cardiac catheterization inspection].

Authors :
Ozaki M
Take T
Sumi S
Ando H
Nakazawa Y
Source :
Nihon Hoshasen Gijutsu Gakkai zasshi [Nihon Hoshasen Gijutsu Gakkai Zasshi] 2011; Vol. 67 (5), pp. 507-16.
Publication Year :
2011

Abstract

We developed software for estimating exposure doses and radiation exposure regions in cardiac catheterization inspection. In the software, the back of a thoracic phantom were divided into a total of 21 square blocks with a width of 30 degrees. Furthermore, we developed a system with which reference air kerma is distributed to each of the above blocks in accordance with the distribution ratio calculated from the data obtained by the system, and the calculation results are displayed. Coronary angiography was performed using thoracic phantoms, and actual measurements were obtained using a fluoroglass dosimeter. The calculated results obtained using the software were compared to the measured results. An almost identical tendency was seen, and the radiation regions of the top three estimated exposure doses were successfully estimated. Radiation region estimation using this software is affected by exposure time as one of its properties. This software enables estimating radiation exposure regions and exposure doses.

Details

Language :
Japanese
ISSN :
1881-4883
Volume :
67
Issue :
5
Database :
MEDLINE
Journal :
Nihon Hoshasen Gijutsu Gakkai zasshi
Publication Type :
Academic Journal
Accession number :
21666374
Full Text :
https://doi.org/10.6009/jjrt.67.507