Back to Search Start Over

Infection or metal hypersensitivity? The diagnostic challenge of failure in metal-on-metal bearings.

Authors :
Galbraith JG
Butler JS
Browne TJ
Mulcahy D
Harty JA
Source :
Acta orthopaedica Belgica [Acta Orthop Belg] 2011 Apr; Vol. 77 (2), pp. 145-51.
Publication Year :
2011

Abstract

The use of second generation metal-on-metal hip articulations has gained favour in the past few years. A hypersensitivity reaction to the metal-on-metal bearing, although rare, is a reported complication and is a novel mode of failure of these implants. Differentiating failure secondary to infection from failure secondary to metal hypersensitivity represents a significant diagnostic challenge. A retrospective review of all cases of hip arthroplasty using metal-on-metal bearings over a 5-year period at a tertiary referral centre identified 3 cases of failure secondary to metal hypersensitivity. Clinical presentation, serological markers, radiological imaging and histological analysis of all cases identified were evaluated. Histological analysis of periprosthetic tissue in all 3 cases identified characteristic features such as perivascular lymphocytic aggregates and chronic inflammation consistent with aseptic lymphocytic vasculitis-associated lesions (ALVAL). This study highlights that failure secondary to metal hypersensitivity must be considered in patients presenting with the reappearance of persistent pain, marked joint effusion, and the development of early osteolysis in the absence of infection.

Details

Language :
English
ISSN :
0001-6462
Volume :
77
Issue :
2
Database :
MEDLINE
Journal :
Acta orthopaedica Belgica
Publication Type :
Academic Journal
Accession number :
21667724