Back to Search
Start Over
Characterization of an extrapolation chamber for low-energy X-rays: experimental and Monte Carlo preliminary results.
- Source :
-
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2012 Jul; Vol. 70 (7), pp. 1388-91. Date of Electronic Publication: 2011 Dec 10. - Publication Year :
- 2012
-
Abstract
- The extrapolation chamber is a parallel-plate ionization chamber that allows variation of its air-cavity volume. In this work, an experimental study and MCNP-4C Monte Carlo code simulations of an ionization chamber designed and constructed at the Calibration Laboratory at IPEN to be used as a secondary dosimetry standard for low-energy X-rays are reported. The results obtained were within the international recommendations, and the simulations showed that the components of the extrapolation chamber may influence its response up to 11.0%.<br /> (Copyright © 2011 Elsevier Ltd. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1872-9800
- Volume :
- 70
- Issue :
- 7
- Database :
- MEDLINE
- Journal :
- Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- Publication Type :
- Academic Journal
- Accession number :
- 22182629
- Full Text :
- https://doi.org/10.1016/j.apradiso.2011.12.012