Transferring calibration coefficients from ionisation chambers used for diagnostic radiology to transmission chambers.

Authors :
Yoshizumi MT
Caldas LV
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2012 Jul; Vol. 70 (7), pp. 1317-20. Date of Electronic Publication: 2011 Dec 26.
Publication Year :
2012

Abstract

In this work, the response of a double volume transmission ionisation chamber, developed at the Instituto de Pesquisas Energéticas e Nucleares, was compared to that of a commercial transmission chamber. Both ionisation chambers were tested in different X-ray beam qualities using secondary standard ionisation chambers as reference dosimeters. These standard ionisation chambers were a parallel-plate and a cylindrical ionisation chambers, used for diagnostic radiology and mammography beam qualities, respectively. The response of both transmission chambers was compared to that of the secondary standard chambers to obtain coefficients of equivalence. These coefficients allow the transmission chambers to be used as reference equipment.<br /> (Copyright © 2011 Elsevier Ltd. All rights reserved.)