Back to Search
Start Over
Refractive index sensor based on a 1D photonic crystal in a microfluidic channel.
- Source :
-
Sensors (Basel, Switzerland) [Sensors (Basel)] 2010; Vol. 10 (3), pp. 2348-58. Date of Electronic Publication: 2010 Mar 22. - Publication Year :
- 2010
-
Abstract
- A refractive index sensor has been fabricated in silicon oxynitride by standard UV lithography and dry etching processes. The refractive index sensor consists of a 1D photonic crystal (PhC) embedded in a microfluidic channel addressed by fiber-terminated planar waveguides. Experimental demonstrations performed with several ethanol solutions ranging from a purity of 96.00% (n = 1.36356) to 95.04% (n = 1.36377) yielded a sensitivity (Δλ/Δn) of 836 nm/RIU and a limit of detection (LOD) of 6 × 10(-5) RIU, which is, however, still one order of magnitude higher than the theoretical lower limit of the limit of detection 1.3 × 10(-) (6) RIU.
Details
- Language :
- English
- ISSN :
- 1424-8220
- Volume :
- 10
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Sensors (Basel, Switzerland)
- Publication Type :
- Academic Journal
- Accession number :
- 22294930
- Full Text :
- https://doi.org/10.3390/s100302348