Traveling-wave ion mobility mass spectrometry of protein complexes: accurate calibrated collision cross-sections of human insulin oligomers.
- Source :
-
Rapid communications in mass spectrometry : RCM [Rapid Commun Mass Spectrom] 2012 May 30; Vol. 26 (10), pp. 1181-93. - Publication Year :
- 2012
-
Abstract
- Rationale: The collision cross-section (Ω) of a protein or protein complex ion can be measured using traveling-wave (T-wave) ion mobility (IM) mass spectrometry (MS) via calibration with compounds of known Ω. The T-wave Ω-values depend strongly on instrument parameters and calibrant selection. Optimization of instrument parameters and calibration standards are crucial for obtaining accurate T-wave Ω-values.<br />Methods: Human insulin and the fast-acting insulin aspart under native-like conditions (ammonium acetate, physiological pH) were analyzed on Waters SYNAPT G1 and G2 HDMS instruments. The calibrated T-wave Ω-values of insulin monomer, dimer, and hexamer ions were measured using many different combinations of denatured and native-like calibrants (masses between 2.85 and 256 kDa) and T-wave conditions. Drift-tube Ω-values were obtained on a modified SYNAPT G1.<br />Results: Insulin T-wave Ω-values were measured at 26 combinations of T-wave velocity and amplitude. Optimal sets of calibrants were identified that yield Ω-values with minimal dependence on T-wave conditions and calibration plots with high R(2)-values. The T-wave Ω-values determined under conditions satisfying these criteria had absolute errors <2%. Structural differences between human insulin and fast-acting insulin aspart were probed with IM-MS. Insulin aspart monomers have increased flexibility, while hexamers are more compact than human insulin.<br />Conclusions: Accurate T-wave Ω-values that are indistinguishable from drift-tube values are obtained when using (1) native-like calibrants with masses that closely bracket that of the analyte, (2) T-wave velocities that maximize the R(2) of the calibration plot for those calibrants, and (3) at least three replicates at T-wave velocities that yield calibration plots with high R(2).<br /> (Copyright © 2012 John Wiley & Sons, Ltd.)
Details
- Language :
- English
- ISSN :
- 1097-0231
- Volume :
- 26
- Issue :
- 10
- Database :
- MEDLINE
- Journal :
- Rapid communications in mass spectrometry : RCM
- Publication Type :
- Academic Journal
- Accession number :
- 22499193
- Full Text :
- https://doi.org/10.1002/rcm.6211