Back to Search
Start Over
Coupling bifurcated stent-grafts to overcome anatomic limitations of endovascular repair of abdominal aortic aneurysms.
- Source :
-
Journal of vascular and interventional radiology : JVIR [J Vasc Interv Radiol] 2012 Aug; Vol. 23 (8), pp. 1065-9. Date of Electronic Publication: 2012 Jun 13. - Publication Year :
- 2012
-
Abstract
- Purpose: To evaluate the effectiveness of the coupling stent-graft technique and outcomes on contrast-enhanced computed tomography (CT).<br />Materials and Methods: All patients who underwent endovascular aneurysm repair (EVAR) for abdominal aortic aneurysms (AAAs) between 2007 and 2010 at a single institution were retrospectively analyzed. Of the 161 cases, 19 patients who had coupling stent-grafting because of AAA proximal neck angulation greater than 60°, conical neck, or iliac tortuosity were included. Patient age ranged from 62 to 87 years (mean, 73.3 y). Mean follow-up was 18.7 months (range, 1-36 mo). The coupling stent-graft technique was defined by the use of suprarenal fixation device main bodies (Zenith or Talent) with one or more EXCLUDER limbs to adapt tortuous and angulated iliac arteries. Pretreatment contrast-enhanced CT was analyzed based on three-dimensional and multiplanar reformatted images. Follow-up contrast-enhanced CT was also analyzed.<br />Results: Mean aneurysm diameter was 68.9 mm ± 14.0, neck length was 32.7 mm ± 15.1, and neck angulation was 68.9° ± 11.5. Conical neck shapes were present in four patients (21.1%) and ruptured AAAs were present in two (10.5%). Among 38 iliac arteries, the mean iliac artery tortuosity index was 1.52 ± 0.27 and iliac angle was 106.8° ± 18.1. Three patients (15.8%) underwent repeat intervention: placement of a balloon-expandable stent in the proximal neck for type I endoleak (n = 1), endovascular embolization for proximal type I endoleak (n = 1), and percutaneous approach following endovascular embolization for type II endoleak (n = 1). No type III endoleak, limb occlusion, limb kinking, or stent-graft migration developed.<br />Conclusions: The coupling stent-graft technique is technically feasible, with acceptable midterm outcomes in EVAR of AAA in patients with unfavorable neck and iliac anatomies.<br /> (Copyright © 2012 SIR. Published by Elsevier Inc. All rights reserved.)
- Subjects :
- Aged
Aged, 80 and over
Aortic Aneurysm, Abdominal diagnostic imaging
Aortography methods
Blood Vessel Prosthesis Implantation adverse effects
Contrast Media
Endovascular Procedures adverse effects
Feasibility Studies
Female
Humans
Iliac Artery diagnostic imaging
Imaging, Three-Dimensional
Male
Middle Aged
Patient Selection
Predictive Value of Tests
Prosthesis Design
Radiographic Image Interpretation, Computer-Assisted
Republic of Korea
Retrospective Studies
Time Factors
Tomography, X-Ray Computed
Treatment Outcome
Aortic Aneurysm, Abdominal surgery
Blood Vessel Prosthesis
Blood Vessel Prosthesis Implantation instrumentation
Endovascular Procedures instrumentation
Stents
Subjects
Details
- Language :
- English
- ISSN :
- 1535-7732
- Volume :
- 23
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- Journal of vascular and interventional radiology : JVIR
- Publication Type :
- Academic Journal
- Accession number :
- 22698969
- Full Text :
- https://doi.org/10.1016/j.jvir.2012.04.014