Back to Search Start Over

Occupational dose in interventional radiology procedures.

Authors :
Chida K
Kaga Y
Haga Y
Kataoka N
Kumasaka E
Meguro T
Zuguchi M
Source :
AJR. American journal of roentgenology [AJR Am J Roentgenol] 2013 Jan; Vol. 200 (1), pp. 138-41.
Publication Year :
2013

Abstract

Objective: Interventional radiology tends to involve long procedures (i.e., long fluoroscopic times). Therefore, radiation protection for interventional radiology staff is an important issue. This study describes the occupational radiation dose for interventional radiology staff, especially nurses, to clarify the present annual dose level for interventional radiology nurses.<br />Materials and Methods: We compared the annual occupational dose (effective dose and dose equivalent) among interventional radiology staff in a hospital where 6606 catheterization procedures are performed annually. The annual occupational doses of 18 physicians, seven nurses, and eight radiologic technologists were recorded using two monitoring badges, one worn over and one under their lead aprons.<br />Results: The annual mean ± SD effective dose (range) to the physicians, nurses, and radiologic technologists using two badges was 3.00 ± 1.50 (0.84-6.17), 1.34 ± 0.55 (0.70-2.20), and 0.60 ± 0.48 (0.02-1.43) mSv/y, respectively. Similarly, the annual mean ± SD dose equivalent range was 19.84 ± 12.45 (7.0-48.5), 4.73 ± 0.72 (3.9-6.2), and 1.30 ± 1.00 (0.2-2.7) mSv/y, respectively. The mean ± SD effective dose for the physicians was 1.02 ± 0.74 and 3.00 ± 1.50 mSv/y for the one- and two-badge methods, respectively (p < 0.001). Similarly, the mean ± SD effective dose for the nurses (p = 0.186) and radiologic technologists (p = 0.726) tended to be lower using the one-badge method.<br />Conclusion: The annual occupational dose for interventional radiology staff was in the order physicians > nurses > radiologic technologists. The occupational dose determined using one badge under the apron was far lower than the dose obtained with two badges in both physicians and nonphysicians. To evaluate the occupational dose correctly, we recommend use of two monitoring badges to evaluate interventional radiology nurses as well as physicians.

Details

Language :
English
ISSN :
1546-3141
Volume :
200
Issue :
1
Database :
MEDLINE
Journal :
AJR. American journal of roentgenology
Publication Type :
Academic Journal
Accession number :
23255753
Full Text :
https://doi.org/10.2214/AJR.11.8455