Back to Search Start Over

Quantification of 5-aminolevulinic acid by CE using dynamic pH junction technique.

Authors :
Hsieh BC
Chen RL
Tsai T
Source :
Journal of separation science [J Sep Sci] 2013 Feb; Vol. 36 (4), pp. 803-8. Date of Electronic Publication: 2013 Jan 22.
Publication Year :
2013

Abstract

An online dynamic pH junction preconcentration method was developed for quantification of 5-aminolevulinic acid (ALA) by CE with the separation time less than 6 min. The optimal dynamic pH junction of ALA was carried out between pH 9.3 borate buffer (BGE, 40 mM) and pH 2.5 phosphate buffer (sample matrix, 40 mM) when 4.1 cm of sample plug was hydrodynamically injected into an uncoated fused-silica capillary (48.5 cm in length, id of 50 μm). If a 24 kV separation voltage was applied, the calibration curve of ALA peak area (200 nm) showed good linearity (R(2) = 0.9991) ranging from 0.01 to 6.5 mg/mL. The reproducibility of this system was excellent with RSDs (n = 10) of 2.5% for peak area response and 0.6% for migration time at ALA concentration of 0.5 mg/mL. The LOD was evaluated as 1.0 μg/mL (S/N > 3). Compared to conventional CE procedure, the sensitivity was successfully improved over 50-fold. The analytical results of pharmaceutical formulations show a good agreement with those by HPLC (r = 0.94).<br /> (© 2013 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.)

Details

Language :
English
ISSN :
1615-9314
Volume :
36
Issue :
4
Database :
MEDLINE
Journal :
Journal of separation science
Publication Type :
Academic Journal
Accession number :
23339045
Full Text :
https://doi.org/10.1002/jssc.201200889