Saliency-guided compressive fluorescence microscopy.

Authors :
Schwartz S
Wong A
Clausi DA
Source :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2012; Vol. 2012, pp. 4365-8.
Publication Year :
2012

Abstract

A novel saliency-guided approach is proposed for improving the acquisition speed of compressive fluorescence microscopy systems. By adaptively optimizing the sampling probability density based on regions of interest instead of the traditional unguided random sampling approach, the proposed saliency-guided compressive fluorescence microscopy approach can achieve high-quality microscopy images using less than half of the number of fluorescence microscopy data measurements required by existing compressive fluorescence microscopy systems to achieve the same level of quality.

Details

Language :
English
ISSN :
2694-0604
Volume :
2012
Database :
MEDLINE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference