Back to Search
Start Over
Intracoronary glycoprotein IIb/IIIa inhibitor infusion via a perfusion coronary catheter to decrease thrombus burden: results from the ClearWay™ Multicenter Registry.
- Source :
-
Cardiovascular revascularization medicine : including molecular interventions [Cardiovasc Revasc Med] 2013 Sep-Oct; Vol. 14 (5), pp. 280-3. Date of Electronic Publication: 2013 Jan 30. - Publication Year :
- 2013
-
Abstract
- Objectives: This multicenter registry aimed to assess the ClearWay™ (CW) perfusion catheter in reduction of thrombus burden and improvement of the coronary flow during percutaneous coronary intervention (PCI).<br />Background: The presence or development of thrombus during PCI is associated with poor prognosis.<br />Methods: The utility of the CW perfusion catheter was assessed in patients who presented with intracoronary thrombus and were subjected to PCI. Data were collected by online survey from 15 US sites. Angiographic assessment of the coronary thrombus burden and the coronary flow after intracoronary infusion of glycoprotein (GP) IIb/IIIa inhibitors via the CW catheter was evaluated at baseline, immediately after infusion, and at the end of the procedure. The cohort included 102 patients; 71.6% presented with ST-elevation myocardial infarction (MI), 21.6% with non-ST-elevation MI, 5.9% with stable angina pectoris, and 2.9% with silent ischemia. The mean cohort age was 59.9±14.5years and comprised mostly of men (72.5%).<br />Results: GP IIb/IIIa inhibitors were infused via the CW catheter on average 1.1±0.3 times, with a mean pressure of 4.2±2.7atm and a mean infusion time of 55±55s. Following the infusion, Thrombolysis In Myocardial Infarction (TIMI) flow improved by 1° in 71 patients (69.6%) and by 2° in 51 patients (50%), while visible thrombus was reduced by 52% (p<0.001). In the final angiogram, TIMI flow was restored in 90.2% and clearance of a visible thrombus was obtained in 91.8% of the lesions.<br />Conclusion: Intracoronary infusion of GP IIb/IIIa inhibitors via the perfusion CW catheter is associated with significant reduction in thrombus burden and with improvement of the coronary flow in patients presenting or developing thrombus burden during PCI.<br /> (Copyright © 2013. Published by Elsevier Inc.)
- Subjects :
- Aged
Coronary Angiography
Coronary Artery Disease blood
Coronary Artery Disease diagnosis
Coronary Artery Disease physiopathology
Coronary Circulation drug effects
Coronary Thrombosis blood
Coronary Thrombosis diagnosis
Coronary Thrombosis physiopathology
Coronary Vessels diagnostic imaging
Coronary Vessels physiopathology
Equipment Design
Female
Health Care Surveys
Humans
Infusions, Intra-Arterial
Male
Middle Aged
Platelet Glycoprotein GPIIb-IIIa Complex metabolism
Registries
Surveys and Questionnaires
Time Factors
Treatment Outcome
United States
Cardiac Catheterization instrumentation
Cardiac Catheters
Coronary Artery Disease therapy
Coronary Thrombosis therapy
Coronary Vessels drug effects
Percutaneous Coronary Intervention adverse effects
Platelet Aggregation Inhibitors administration & dosage
Platelet Glycoprotein GPIIb-IIIa Complex antagonists & inhibitors
Subjects
Details
- Language :
- English
- ISSN :
- 1878-0938
- Volume :
- 14
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Cardiovascular revascularization medicine : including molecular interventions
- Publication Type :
- Academic Journal
- Accession number :
- 23375788
- Full Text :
- https://doi.org/10.1016/j.carrev.2012.12.006